Abbexa's Cell-Free DNA Kit is an ideal solution for lysing samples by enzyme hydrolysis and purifying cell-free DNA by specific adsorption of silica magnetic beads. It is suitable for isolating and purifyling high quality cell-free DNA from 0.5-10 ml serum or plasma. The extracted DNA can be used in PCR, qPCR and NGS, and also magnetic beads-based nucleic acid extraction systems. Kit Components:
Binding Buffer: 120 ml
Clean Buffer: 30 ml
Wash Buffer: 24 ml
Elution Buffer: 4 ml
Proteinase K (20 mg/ml): 3 × 1 ml
20% SDS: 6 ml
Magnetic Cell-Free Beads: 2 ml
Material Required But Not Provided:
Magnetic Stand (with tube holder)
Vortex Mixer
Centrifuge and Microcentrifuge Tubes
Pipettes and Pipette Tips
Isopropyl Alcohol
Target
Cell-Free DNA
Storage
Store magnetic beads at 2-8°C. Do not freeze. Store all other reagents at room temperature.
Shelf Life
12 months.
Availability
Shipped within 10-20 working days.
Note
THIS PRODUCT IS FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC, THERAPEUTIC OR COSMETIC PROCEDURES. NOT FOR HUMAN OR ANIMAL CONSUMPTION.
Directions for use
Reagent Preparation:
Working Binding Buffer: Add 40 ml isopropryl alcohol to 120 ml Binding Buffer to prepare the Working Binding Buffer solution.
Working Clean Buffer: Add 30 ml isopropryl alcohol to 30 ml Clean Buffer to prepare the Working Clean Buffer solution.
Working Wash Buffer: Add 96 ml isopropryl alcohol to 24 ml Wash Buffer to prepare the Working Wash Buffer solution.
Notes:
Vortex magnetic beads before use.
Use germ-free, nucleic-acid-free and nuclease-free centrifuge tubes and pipette tips.
The component content of cell-free DNA is extremely low, we recommend using low nucleic acid adsorption centrifugal tubes for plasma storage, DNA extraction and DNA storage.
Avoid repeated freeze/thaw cycles of reagents, samples and isolated DNA.
Reaction System: Prepare the reaction system in a 15 ml or 50 ml centrifuge tube according to the table below.
Component
Plasma Volume
0.5 ml
1 ml
2 ml
4 ml
10 ml
20% SDS
25 µl
50 µl
100 µl
200 µl
500 µl
Proteinase K
15 µl
30 µl
60 µl
120 µl
300 µl
Working Binding Buffer
0.75 ml
1.5 ml
3 ml
6 ml
15 ml
Magnetic Cell-Free Beads
10 µl
20 µl
40 µl
80 µl
200 µl
Assay Procedure:
Set up a reaction system according to the table above.
Vortex the tubes for 15 seconds, then leave at room temperature for 20 minutes. Invert the tube 3-5 times throughout this time period.
Magnetic separation: Place the centrifuge tubes on a magnetic stand, then gently spin the tubes left and right by hand. Reverse the spin direction when the magnetic beads begin to aggregate towards the tube wall, and repeat 2-3 times. Ensure that any beads on the lid aggregate to the tube wall. Allow to stand for 2 minutes and ensure all beads are aggregated to the tube wall.
Discard the supernatant from the opposite side of the magnetic beads, taking care not to remove the beads themselves. Take the tubes off the magnetic stand and add 1 ml of Working Clean Buffer (with isopropryl alcohol). Vortex for 15 seconds, then transfer to a new 1.5 ml centrifuge tube. If magnetic beads remain in the original tube, transfer the supernatant back to the original tube, wash, then transfer to the 1.5 ml centrifuge tube. Repeat Step 3 to carry out another magnetic separation.
Discard the supernatant. Take the tubes off the magnetic stand and add 1 ml of Working Wash Buffer solution (with isopropyl alcohol). Vortex for 15 seconds, then repeat Step 3 to carry out another magnetic separation.
Repeat Step 5.
Discard the supernatant, including any liquid on the lid. It is recommended to use smaller size pipette tips to remove the supernatant thoroughly.
Allow to stand and air-dry for 5-10 minutes.
Add Elution Buffer according to the table below
Component
Original Plasma Volume
0.5 ml
1 ml
2 ml
4 ml
10 ml
Elution Buffer
20 µl
30 µl
50 µl
75 µl
200 µl
Vortex the tubes for 5 minutes.
Place the centrifuge tubes on the magnetic stand. Transfer the liquid into a new 1.5 ml low nucleic acid adsorption centrifuge tube, taking care not to remove the beads themselves.
We use cookies necessary for the function of our website, to improve your experience, analyse our traffic, and cookies that optimise the performance. To learn more about our cookies, please read our Cookie Notice.